Microscopic Laboratory
Laboratory Manager: Eng. Khosravi
Email: M.khosravi@ippi.ac.ir - Sem@ippi.ac.ir
Contact Number: 021-48662135
Microscopic Laboratory
Currently, various methods exist for identifying and analyzing materials, one of the most famous being microscopic methods. Using microscopic methods, high-magnification images of the material are obtained to study its details with greater accuracy. Two parameters, resolution and depth of field, are very important and influential in microscopic images, especially at high magnifications. Both parameters are highly dependent on the wavelength of light used in the microscope. In optical microscopes, image magnification can be increased to any extent by changing the curvature of the lenses (degree of concavity and convexity) and their number. However, due to the long wavelength of light, images practically lose resolution at magnifications above 2000. Therefore, in electron microscopes, electrons are used instead of light due to their short wavelength, enabling the acquisition of images at higher magnifications.
Microscopic Laboratory Equipment
Vega2 SEM Scanning Electron Microscope manufactured by Tescan
Technical Specifications:
Magnification range: 20X-100,000X
Voltage range: 5kV-30kV
Vacuum level: 10-5 Pa
Sample Conditions: Samples must be in powder or solid form, stable in vacuum, free of any moisture and volatile substances, indestructible in vacuum, non-explosive, non-radioactive, non-allergenic, or emitting any other harmful radiation.
Applicable Standard: (ASTM: Standard Practice for Characterization of
Particles (F1877-05))
The microscope is equipped with a Back Scattered/Electron (BSE) detector and provides the following capabilities:
- Imaging of phases on surfaces (if the average atomic number of the phases differs from the matrix).
- Grain boundary observation in unetched samples.
- Evaluation of crystallographic orientation of grains with diameters as small as 2 to 10 micrometers.
- Chemical analysis EDS model INCA manufactured by Oxford Instruments
Technical Specifications:
Voltage range: 5kV-30kV
Element detection capability from Carbon upwards
Detection limit for minor elements down to 0.1 wt%
Analysis capabilities: Point, line, and mapping. Sample Conditions: Samples must be in powder or solid form, stable in vacuum, free of any moisture and volatile substances, indestructible in vacuum, non-explosive, non-radioactive, non-allergenic, or emitting any other harmful radiation.
Applicable Standard:
ASTM: Standard Guide for Quantitative Analysis by Energy-
Dispersive Spectroscopy (E 1508-98)
Sputter Coating Deposition Systems
1. Model K450X manufactured by EMITECH
2. Model DSR1 manufactured by Nano Structured Coatings
For preparing non-conductive samples for SEM testing, sample coating is available up to a thickness of 15 nm.
Current range: 5 mA-25 mA
Vacuum level: 10-1 Pa
Sample Conditions: Samples must be in powder or solid form, stable in vacuum, free of any moisture and volatile substances, indestructible in vacuum, non-explosive, non-radioactive, non-allergenic, or emitting any other harmful radiation.
Reflected Light Optical Microscope Model JENAVERT manufactured by Carl Zeiss
Technical Specifications:
Lamp: Visible light
Lenses: 50X-100X-200X-500X-1000X
Equipped with digital camera
Sample Preparation and Mounting: The sample is glued and fixed on a flat surface according to its application (surface study, cross-section, or particle size distribution, etc.). This assembly is placed on a movable and rotatable flat stage.
Transmitted Light Optical Microscope Model JENAPOL manufactured by Carl Zeiss
Technical Specifications:
Lamp: Visible light
Lenses: 50X-100X-200X
Equipped with two Polaroid filters
Equipped with digital camera
For determining dimensions and morphology, samples must be transparent or semi-transparent and prepared on glass slides.
Transmitted Light Optical Microscope Model JENAPOL with Hot Stage manufactured by Carl Zeiss
Technical Specifications:
Lamp: Visible light
Lenses: 100X
Equipped with two Polaroid filters
Temperature range: Ambient to 2500C
Equipped with digital camera
Samples must be completely transparent and thin. For this purpose, microtomy, hot pressing, or solvent casting methods can be used for sample preparation.
Atomic Force Microscope (AFM) Model E 95-200 by DME
Sample in solid or powder form, completely dry and free of any dust or fingerprints.
Sample surface must be completely flat (roughness less than 15µm), sample dimensions less than 3cm×3cm.