Dynamic-Mechanical Analysis Laboratory
Laboratory Director: Masoumeh Masoudi
Contact Number: 021-48662145
Email: DMTA@ippi.ac.ir m.masoudi@ippi.ac.ir ,
Dynamic-Mechanical Analysis Laboratory
(Dynamic-Mechanical Analysis Laboratory)
To complete its analytical equipment systems, the Polymer and Petrochemical Research Institute of Iran has equipped the Dynamic-Mechanical Analysis Laboratory with DMA-Triton and DMA Mettler Toledo instruments. In simple terms, DMA analysis involves applying an oscillatory force to a sample and analyzing the sample's response to it. The Storage Modulus, Loss Modulus, and Damping Factor (tan δ) of polymeric materials, which arise from their viscoelastic properties, are examined in this test. In this test, the sample is subjected to periodic stresses at a specified frequency (standard 1 Hz) in various modes such as bending, tensile, shear, and compression, under isothermal or non-isothermal conditions, with very low strain so that the material remains in its linear viscoelastic region. Changes in Storage Modulus, Loss Modulus, and the Damping Factor (tan δ), which is the ratio of Loss Modulus to Storage Modulus, are measured as a function of temperature. Using this test, transitions such as α, β, γ, etc., which result from various motions originating from the main chain or side groups attached to it, are observed. Additionally, the glass transition temperature (Tg), which is the temperature at which the sample's Storage Modulus drops suddenly and the Damping Factor increases sharply, can also be calculated and analyzed by this technique.
Type, Shape, and Dimensions of Samples:
Test Mode | Thickness (mm) | Width (mm) | Length (mm) | Number of Samples | Example | |
Tensile | <1 | 5-8 | 35 | 3 | Film | |
Bending | Single/Dual | 1-2 | 5-8 | 35 | 3 | Plastic / Rubber |
3 point | 2-3 | 5-8 | 35 | 3 | Composite | |
Compression | To be coordinated with the laboratory | 5-8 | 5-8 | 3 | Foam / Gel | |
Shear | <2 | 7-8 | 7-8 | 6 | Hydrogel, Adhesive, etc. | |
· Test Conditions
v Test Environment: Air
v Achievable Temperature Range: -150 to 250 degrees Celsius
v Frequency Range: up to 180 Hz
v Heating rate or frequency step is determined by the customer.
v Samples must never melt or flow at the test temperature.
v Any irregularities or inconsistencies in sample dimensions will cause errors in the test results.
v Powdered samples can also be used to determine Tg.
Standards Used:
ASTM E1640: Standard Test Method for Assignment of the Glass Transition Temperature by Dynamic Mechanical Analysis
ASTM D5023: Standard Test Method for Plastics: Dynamic Mechanical Properties: in Flexure (Three Point Bending)
ASTM D5026: Standard Test Method for Plastics: Dynamic Mechanical Properties: in Tension
Laboratory Equipment and Instruments:
· DMA-Mettler Toledo (DMA1, Made in Switzerland)
· DMA-Triton (Tritec 2000, Made in England)